Abstract

Dynamic susceptibility contrast (DSC) MRI is now increasingly used for measuring perfusion in many different applications. The quantification of DSC data requires the measurement of the arterial input function (AIF) and the deconvolution of the tissue concentration time curve. One of the most accepted deconvolution methods is the use of singular value decomposition (SVD). Simulations were performed to evaluate the effects on DSC quantification of the presence of delay and dispersion in the estimated AIF. Both delay and dispersion were found to introduce significant underestimation of cerebral blood flow (CBF) and overestimation of mean transit time (MTT). While the error introduced by the delay can be corrected by using the information of the arrival time of the bolus, the correction for the dispersion is less straightforward and requires a model for the vasculature.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.