Abstract

Aim: To compare the difference between the measured and target values in vancomycin external quality assessment (EQA) samples and to investigate the factors for underestimating its concentration. Materials & methods: A retrospective analysis of 195 international vancomycin EQA results was performed. Deviations of the concentrations determined by TDx fluorescence polarization immunoassay (FPIA), Axsym FPIA and Architect chemiluminescence microparticle immunoassay (CMIA)method were -2.43, -16.28and -10.53%, respectively. Chromatographic peaks of the crystalling degradation products appeared in samples with large deviations. Vancomycin were degraded after long-term transporting and high temperature. Conclusion: Vancomycin concentrations measured by Axsym FPIA and Architect CMIA methods were likely to be underestimated. Long-term transporting resulted in low EQA results, suggesting that establishing a local EQA system for vancomycin is essential.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.