Abstract

Frosting is a common phenomenon in refrigeration and frozen equipment, which degrades heat transfer performance and leads to more energy consumption. This paper investigates the application of hot gas bypass defrosting as an alternative to the electric heater defrosting method. The hot gas from the discharge port enters the evaporator to melt the frost layer on the surface of the evaporator. Experimental data indicate that more than 80% defrosting efficiency can be achieved. The defrosting efficiency increases with thicker frost layer and larger bypassed flow rate. In the end, a defrosting control strategy dedicated to medical refrigerator application is proposed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.