Abstract

In this research, effects of material properties of stent materials, such as nitinol and ferrous alloy, on a self-expanding stent are studied by the stent deployment simulations using finite element method (FEM). Stent, vessel, and plaque are modeled by using AutoCAD. The model is imported to a finite element analysis software Ansys 15.0 and the deployment analysis of the stent is carried out. In addition, two methods of the unsheathing of the self-expanding stent are studied: (1) a conventional method where a stent comes out from one end of the sheath and (2) a new method in which a stent comes out from the center of the sheath. Also, both the area stenosis ratio and Dog-Boning are discussed referring the simulations conducted in the paper.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.