Abstract

The National Institute of Standards and Technology has designed and built two electrically calibrated laser calorimeters as primary standards for absolute energy measurements at the wavelength of 248 nm. Under the sponsorship of SEMATECH, NIST developed the calorimeters to improve measurement of dose energy in excimer laser based microlithography. The calorimeter system can be used to calibrate transfer standards which in turn can be used to calibrate detectors employed for energy measurements of semiconductor wafer exposure. The excimer calorimeter uses a glass filter which functions as a volume absorber that allows collection of nanosecond pulses of laser radiation without suffering damage. The measurement range of the calorimeters is 0.3-25 J, but can be extended to 1 mJ with beamsplitters. Electrical calibration of the calorimeters shows a standard deviation in the calibration factor of less than 0.5% for entire energy range. The total uncertainty of typical power and energy meter calibrations is approximately 2%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.