Abstract
AbstractThe uncertainty of deep ocean temperature (~1°C) measurement was evaluated. The time drifts of six deep ocean standards thermometers were examined based on laboratory calibrations as performed by the manufacturer in triple point of water (TPW) cells and gallium-melting-point (GaMP) cells. The time drifts ranged from −0.11 to 0.14 mK yr−1. Three of the six thermometers were evaluated at the National Metrology Institute of Japan in five TPW cells and a GaMP cell, and the temperature readings agreed with the realized temperature of the national standard cells of Japan within ±0.14 and ±0.41 mK for TPW and GaMP, respectively. The pressure sensitivities of the deep ocean standards thermometers were estimated by comparison with conductivity–temperature–depth (CTD) thermometers in the deep ocean, and no notable difference was detected. Pressure sensitivities of the two CTD thermometers were examined by laboratory tests, and the results suggest that the deep ocean standards thermometers have no pressure sensitivity, at least up to 65 MPa. The position and attitude motion of the CTD system can affect temperature and salinity data quality. The overall expanded uncertainty of the deep ocean temperature measurement (up to 65 MPa) by the CTD thermometer calibrated in reference to the deep ocean standards thermometer is estimated to be 0.7 mK.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.