Abstract

We report a case of deep brain stimulation (DBS) hardware failure due to emergently performed subcutaneous coronary angioplasties complicated by cardioversion for rapid worsening of angina pectoris and some trouble shooting problems emerged after invasive cardiovascular procedures. The patient with prior implantation of permanent pacemaker due to vasovagal syndrome underwent successful left-sided unilateral electrode implantation into the subthalamic nucleus. During 21 months follow-up period the patient experienced 2 times episodes of aggravation of unstable angina pectoris 15 and 21 months respectively, which necessities emergent coronary angioplasties. After the first emergently performed coronary angioplasty with cardioversion the interrogation of DBS system revealed the depletion of an internal pulse generator (IPG). The secondly performed coronary angioplasty complicated by ventricular tachyarrhythmia with DBS system switched on during emergent cardioversion resulted in partial dysfunction of DBS electrode. Patients harboring cardiovascular implantable electronic devices (CIEDs) and DBS systems require special attention and good cooperation of neurosurgeons, interventional cardiologist, and neurologist. Some emergently performed invasive cardiovascular procedures which necessities cardioversion may cause DBS hardware failure with subsequent worsening of movement disorder symptoms.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.