Abstract
Cone-beam dedicated breast CT scanners have been developed to provide 3D images of the breast with sub millimeter isotropic resolution, no breast compression, and dose equivalent to mammography. This article will first review the system design and hardware components used in dedicated breast CT, followed by a review of scanner performance and image quality investigations. Proposed reconstruction algorithms will be discussed and the results from clinical trials presented. Promising new directions in breast CT, such as multi-modality, volume of interest, and multi-energy imaging, will be described. The paper will also discuss the challenges of breast CT including anatomical coverage and image management. Keywords: CT, cone-beam, breast imaging
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.