Abstract

Instruments based on light obscuration sensors are widely used for measuring the size distribution of insoluble sub-visible particles in liquid suspensions, being fast and suitable for in situ and real-time measurements. Such instruments are typically calibrated by means of reference polystyrene spherical particles with a specific refractive index, which unavoidably leads to systematic errors when determining the size of particles of different materials. In this paper, we propose a reliable and consistent method to overcome this limitation by setting the refractive index value according to the sample, thus achieving an improved particle size distribution (PSD) measurement. An ad hoc, ready-to-use, open source code with a graphical interface able to drive an in-line instrument and obtain a real-time correction to the PSD has been developed. The method has been extensively validated with several oil emulsions characterized by different refractive index values and the results have been compared with an independent optical method. As an example of application, we have adopted this approach for the analysis of dust suspended in meltwater of an ice core from a glacier in the Aosta Valley (Italy). We believe that our approach will strongly improve the accuracy in characterizing liquid suspensions and reduce discrepancies between data obtained with different methods. The code has been made publicly available at: https://instrumentaloptics.fisica.unimi.it/dedalo/ and on the GitHub page of the corresponding author (https://github.com/LucaTeruzzi/DEDALO).