Abstract

Use of non-invasive ventilation (NIV) in very low birthweight infants to decrease the incidence of bronchopulmonary dysplasia can also lead to pressure injuries (PI) caused by the respiratory device interface. We aimed to decrease our incidence of PIs related to the mask/prongs interface used for NIV (PI-NIV). We identified correct use of barriers and appropriate interface fit as key targets for intervention. Over several PDSA cycles, we developed custom 3D printed barrier templates to allow for barriers to be cut at the bedside and created concise educational documents to assist with interface fitting and troubleshooting. The incidence of all PI-NIV decreased from 5.64 to 2.27 per 1000 NIV patient-days and the incidence of reportable (stage 3-4 and unstageable) PI-NIV decreased from 1.13 to 0 per 1000 NIV patient-days during the study period. With appropriate barrier usage and targeted education, the risk of PI-NIV can be minimized.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.