Abstract

Capillary electrophoresis (CE) can be used for the rapid determination of pharmaceuticals, particularly in routine quality control analysis. This paper focuses on several approaches aimed at decreasing the analysis time with commercially available instrumentation by (i) application of a high electric field through a reduced capillary, (ii) use of a dynamically coated capillary to increase the electroosmotic flow, (iii) short-end injection (SEI) technique, and (iv) application of multiple sample injections. Moreover, SEIs were combined with the three other approaches. A pharmaceutical formulation containing lidocaine as an active component was selected, and the methods were validated according to the ICH guidelines. The seven approaches investigated fulfilled different statistical requirements and demonstrated their linearity and trueness, with good recoveries and confidence limits always inferior to 1.5%. Furthermore, relative standard deviation (RSD) values for repeatability and intermediate precision were inferior to 1.1 and 1.8%, respectively. These results confirmed that each approach is of utmost interest to increase the analyte throughput in CE.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.