Abstract

To compare rates of vitreous loss during cataract surgery when a silicone-tipped irrigation/aspiration (I/A) instrument is used for cortex removal as opposed to a metal tip. Parkland Memorial Hospital, Dallas, Texas, USA. A retrospective chart review of all patients who had cataract extraction by phacoemulsification by third-year ophthalmology residents between September 2000 and February 2004 was conducted. A silicone-tipped I/A instrument was used to remove cortex for all surgeries beginning in September 2002, whereas a metal I/A tip was used previously. Of the 1072 cases performed with a metal I/A tip, there were 13 cases of vitreous loss during cortex removal (rate 1.2%) and 26% of all vitreous loss during that time occurred during cortex removal. Of the 805 cases performed with a silicone I/A tip, there was a single case (0.1%) of vitreous loss during cortex removal (P=.004); only 4% of all vitreous loss during that time occurred during cortex removal (P=.011). The overall incidence of vitreous loss during cortex removal and the ratio of vitreous loss during cortex removal to all cases of vitreous loss were significantly decreased using the silicone-tipped I/A instrument.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.