Abstract
In this work, 213Bi has been separated from an open 225Ac source by collecting recoil atoms onto a glass plate in vacuum. The activity of such recoil sources has been measured as a function of time, using an ion-implanted planar Si detector in quasi-2π geometry. From these measurements, a new half-life value of T1/2(213Bi)=45.62 (6)min was derived. Additionally, high-resolution alpha-spectrometry measurements were performed at a solid angle of 0.4% of 4πsr, to verify the energies and emission probabilities of the α-emissions from 213Bi. Using 225Ac, 221Fr, 217At and 213Po peaks as reference peaks, the measured 213Bi α-peak energies at Eα,0=5878 (4)keV and Eα,1=5560 (4)keV were about 10keV higher than validated data. The relative α-particle emission probabilities of 213Bi, Pα,0=0.9155 (11) and Pα,1=0.0845 (11), and the 213Bi alpha branching factor, Pα=1−Pβ=2.140 (10)%, are compatible with recommended values, but have a higher accuracy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.