After the transfer of the thick copper target plate irradiated with 3 meV protons from a medical cyclotron, the gamma dose rate in the target cave was monitored every minute for 25 hours using wall-mounted gamma area monitors. The dose rate decay curve was fitted with four exponential functions. By analysing the slope of the exponentials the traces of radioactive 28Al, 56Mn 24Na and 59Fe were identified. The results were used to minimize personnel radiation exposure during maintenance work on the cyclotron.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.