Abstract

We report the death of a young man during a laparoscopic partial splenectomy performed with an argon plasma coagulator to remove a benign cyst. The report analyzes the very particular mechanism of a gas embolism, which caused death here. This analysis leads us to recommend a close attention on the use of argon coagulators during laparoscopy. The aim of this article is to draw surgeons' attention to the conclusions of a court-ordered expert assessment intended to elucidate the mechanisms responsible for the death of a 20-year-old man during a laparoscopic partial splenectomy performed with an argon plasma coagulator to remove a benign cyst.