Abstract

The characteristics of the dead zone of a gas counter were studied at gas pressures of 760-150 torr of Ar(75)+isoC/sub 4/H/sub 10/(25) mixture and compared with those of a commercially available GM tube. The counter operated in the SQS (self-quenching streamer) mode at pressures higher than 350 torr, and in the GM mode at the lower pressures; the operation mode changed around 350 torr, where the mean free path of VUV (vacuum ultraviolet) photons is about 50 mu m. The dead zone of the counter used was evaluated to be 290 mu s-cm for the SQS mode (at 760 torr) and 2450 mu s-cm for the GM mode (at 150 torr). The counting rate capability of the SQS tube is more than one order of magnitude larger than that of the GM tube.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.