Abstract

We have developed and investigated the main parts of a voltage calibrator with an output voltage up to 10 V. The calibrator is based on a quantum DC voltage which is generated by an array of high-temperature superconductor Josephson junctions. The described concept and the demonstrated parameters of the main parts of the instrument show, that a combined uncertainty of the output voltage of a few parts in 10/sup 8/ can be obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.