This paper presents a comparison results between the Quantum Hall System (QHS) under development at the Quantum Electrical Metrology Laboratory (Lameq) and the current comparator calibration system, traceable to the Bureau International des Poids et Mesures (BIPM), at the Electrical Standardization Metrology Laboratory (Lampe), both part of the Electrical Metrology Division, at Inmetro. Comparisons were performed with 1 Ω, 10 Ω, 100 Ω, 1 kΩ and 10 kΩ resistors. The results obtained over two years of work are presented here, showing that the comparison contributed to improve the calibration systems of both Lampe and Lameq.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.