Abstract

Clinically feasible assessment of self-feeding is important for adults and children with motor impairments such as stroke or cerebral palsy. However, no validated assessment tool for self-feeding kinematics exists. This work presents an initial validation of an instrumented spoon (DataSpoon) developed as an evaluation tool for self-feeding kinematics. Ten young, healthy adults (three male; age 27.2 ± 6.6 years) used DataSpoon at three movement speeds (slow, comfortable, fast) and with three different grips: “natural”, power and rotated power grip. Movement kinematics were recorded concurrently using DataSpoon and a magnetic motion capture system (trakSTAR). Eating events were automatically identified for both systems and kinematic measures were extracted from yaw, pitch and roll (YPR) data as well as from acceleration and tangential velocity profiles. Two-way, mixed model Intraclass correlation coefficients (ICC) and 95% limits of agreement (LOA) were computed to determine agreement between the systems for each kinematic variable. Most variables demonstrated fair to excellent agreement. Agreement for measures of duration, pitch and roll exceeded 0.8 (excellent agreement) for >80% of speed and grip conditions, whereas lower agreement (ICC < 0.46) was measured for tangential velocity and acceleration. A bias of 0.01–0.07 s (95% LOA [−0.54, 0.53] to [−0.63, 0.48]) was calculated for measures of duration. DataSpoon enables automatic detection of self-feeding using simple, affordable movement sensors. Using movement kinematics, variables associated with self-feeding can be identified and aid clinical reasoning for adults and children with motor impairments.

Highlights

  • Recent technological advances have enabled the development of lightweight, wearable inertial motion sensors, which are showing promise as rehabilitation tools [1]

  • We present the initial validation of an instrumented spoon (DataSpoon) [13,14], developed as an assessment tool for clinicians that provides quantitative information regarding self-feeding in children and adults with motor impairments such as cerebral palsy (CP) or stroke

  • The present results indicate that for most kinematic measures, concurrent validity of movement quality measures, which were extracted automatically from both systems, was fair to excellent when evaluated for young, healthy individuals

Read more

Summary

Introduction

Recent technological advances have enabled the development of lightweight, wearable inertial motion sensors, which are showing promise as rehabilitation tools [1]. Sensor-based assessment of movement kinematics is currently used for gait analysis in healthy individuals [3] as well as clinical populations such as Parkinson’s disease [4], stroke and Huntington’s. Upper limb kinematics have been recorded using wearable sensors in healthy individuals [7] as well as individuals after stroke [8], in order to objectively quantify movement patterns. Inertial sensors are able to detect performance of functional tasks such as drinking or brushing hair, in both healthy and clinical populations [9]. Information derived from low-cost inertial sensors may be able to be used to provide feedback (e.g., auditory, visual, tactile), and affect motor performance as is currently the case with more high-end motion capture systems [10]

Methods
Results
Discussion
Conclusion