Abstract

X-ray absorption spectroscopy on dilute samples is still a challenge nowadays. By means of the fluorescence processes that follow the absorption it is possible to obtain the XAS spectrum; however, in order to gain sufficient statistics a multi-pixel fluorescence detector is required. A 100 pixel Ge detector will be available at beamline P64 at PETRA III for ultra dilute sample systems. Here, we propose a method of data reduction that resolves the line of interest from the scattering peak, improving the signal-to-noise ratio of the final XAS spectrum compared with the one obtained with a simple region of interest (ROI).

Highlights

  • Chemical processes on biological samples, like metalloproteins, are of great interest

  • The fitting algorithm was automatized to fit the amplitude of the Gaussian Agauss corresponde√nt to the CuKα line at each incident energy and to determine the area under the Gaussian ( 2πAgaussσgauss) in order to build up the XANES spectrum

  • Here we present an automatic routine of data reduction for the 100 pixel Ge detector with X-ray Absorption Spectroscopy (XAS) experiments

Read more

Summary

Introduction

Chemical processes on biological samples, like metalloproteins, are of great interest. 2. Data Reduction Reducing n fluorescence spectra at a single incident energy could be done in principle with an average. A comparison between the three methods is exhibited in Figure 2 and shows that the corrected mean and the median decrease slightly the overlapping between the line of interest and the scattering.