Abstract
Crossed-strip gamma-ray detectors are an attractive option for small-animal SPECT imagers due to their high space-bandwidth product. In systems with independent triggering of the two sides of the detector, advanced data-processing techniques are required to accurately determine gamma-ray interaction locations and energy deposition. Optimal detector operation further relies on rigorous detector characterization in order to achieve detector triggering uniformity and best timing resolution and to permit position and energy estimation with maximum-likelihood methods. We describe algorithms and methods developed for calibrating and characterizing a recently fabricated system based on 1024-strips-per-side 1-mm-thick silicon detectors.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Nuclear Science Symposium conference record. Nuclear Science Symposium
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.