Abstract

Natural olfaction suggests that numerous replicas of small sensors can achieve large sensitivity. This concept of sensor redundancy can be exploited by use of optical chemical sensors whose use of image sensors enables the simultaneous measurement of several spatially distributed indicators. Digital image sensors split the framed scene into hundreds of thousands of pixels each corresponding to a portion of the sensing layer. The signal from each pixel can be regarded as an independent sensor, which leads to a highly redundant sensor array. Such redundancy can eventually be exploited to increase the signal-to-noise ratio. In this paper we report an algorithm for reduction of the noise of pixel signals. For this purpose, the algorithm processes the output of groups of pixels whose signals share the same time behavior, as is the case for signals related to the same indicator. To define these groups of pixels, unsupervised clustering, based on classification of the indicator colors, is proposed here. This approach to signal processing is tested in experiments on the chemical sensitivity of replicas of eight indicators spotted on to a plastic substrate. Results show that the groups of pixels can be defined independently of the geometrical arrangement of the sensing spots, and substantial improvement of the signal-to-noise ratio is obtained, enabling the detection of volatile compounds at any location on the distributed sensing layer.

Highlights

  • Observation of changes of the optical properties of sensing molecules provides a powerful method of detection of the occurrence of molecular recognition [1]. This is complemented by optoelectronics that provides a number of devices enabling assembly of sensors that transduce the changes of optical properties into electric signals

  • Image sensors offer a simple way of developing large sensor arrays to study the behavior of a number of replicas of the same chemical reporter [6] or even to signal the diffusion of gases into polymers [7]

  • The tested chemical indicators were six metal complexes of 5,10,15,20-tetraphenylporphyrin, namely (5,10,15,20-tetraphenylporphyrin)palladium [PdTPP], (5,10,15,20-tetraphenylporphyrin)rhodium [RhTPP], (5,10,15,20-tetraphenylporphyrin) zinc [ZnTPP], (5,10,15,20-tetraphenylporphyrin)molybdenum [MoTPP], (5,10,15,20-tetraphenylporphyrin)iron [FeTPP], and (5,10,15,20-tetraphenylporphyrin)manganese [MnTPP], a silicon(phthalocyanine) [SiPC], and a pH indicator known as Nile blue (NB)

Read more

Summary

ORIGINAL PAPER

Data processing for image-based chemical sensors: unsupervised region of interest selection and background noise compensation. Francesca Dini & Eugenio Martinelli & Roberto Paolesse & Daniel Filippini & Detlev Schild & Ingemar Lundström & Corrado DI Natale. Received: 6 August 2011 / Revised: 16 October 2011 / Accepted: 18 October 2011 / Published online: 7 November 2011 # Springer-Verlag 2011

Introduction
Sensing layer
Measurement setup
Processed data
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call