Abstract

A basic assumption in the classic computed tomography (CT) theory is that an object remains stationary in an entire scan. In biomedical CT/micro-CT, this assumption is often violated. To produce high-resolution images, such as for our recently proposed clinical micro-CT (CMCT) prototype, it is desirable to develop a precise motion estimation and image reconstruction scheme. In this paper, we first extend the Helgason-Ludwig consistency condition (HLCC) from parallel-beam to fan-beam geometry when an object is subject to a translation. Then, we propose a novel method to estimate the motion parameters only from sinograms based on the HLCC. To reconstruct the moving object, we formulate two generalized fan-beam reconstruction methods, which are in filtered backprojection and backprojection filtering formats, respectively. Furthermore, we present numerical simulation results to show that our approach is accurate and robust.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.