Abstract

The random nature of self-amplified spontaneous emission (SASE) is a well-known challenge for x-ray core level spectroscopy at SASE free-electron lasers (FELs). Especially in time-resolved experiments that require a combination of good temporal and spectral resolution the jitter and drifts in the spectral characteristics, relative arrival time as well as power fluctuations can smear out spectral-temporal features. We present a combination of methods for the analysis of time-resolved photoelectron spectra based on power and time corrections as well as self-referencing of a strong photoelectron line. Based on sulfur 2p photoelectron spectra of 2-thiouracil taken at the SASE FEL FLASH2, we show that it is possible to correct for some of the photon energy drift and jitter even when reliable shot-to-shot photon energy data is not available. The quality of pump–probe difference spectra improves as random jumps in energy between delay points reduce significantly. The data analysis allows to identify coherent oscillations of 1 eV shift on the mean photoelectron line of 4 eV width with an error of less than 0.1 eV.