Abstract

Steady-state free precession cine images from cardiac magnetic resonance imaging studies of 24 patients were reviewed retrospectively to identify dark flow artifacts. The cause and features of the artifacts were studied in flow phantom experiments. Dark flow artifacts were recognized in eight of the 24 cases and were characterized by low or inhomogeneous signal intensity in blood pools with little change in adjacent tissues. The artifacts could be mimicked in flow phantom experiments by deliberately deshimming the gradients and appeared periodically during imaging with off-centered frequencies. These artifacts appeared to be caused by spins moving within an inhomogeneous magnetic field.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.