Abstract
The Periotest device was used to evaluate the damping characteristics of different bone-to-implant interfaces of Brånemark implants. Implant and abutment lengths were found to have a significant influence on Periotest value (PTV). There was no statistical difference between PTVs of implants located in the anterior and posterior areas of the same jaw at abutment connection. No differences were found between PTVs of standard and self-tapping implants. PTVs at abutment connection were highly influenced by the bone quality. The contact with two corticals, one at each extremity of the implant, was of a greater influence in the maxillary than in the mandibular jaw, indicating a firmer anchorage between the bi-cortical implants and the surrounding bone. A progressive decrease of PTVs over time was observed, significant up to 5 years of loading.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.