Abstract

Two series of present-day pacemakers were tested in vitro with pulsed x-ray radiation. The first series of 12 pacemakers consisted of 10 different types and models of demand pacemakers (VVI). The second series of 13 pacemakers had 9 different types and models of programmable pacemakers. Unlike the first series which showed only mild changes in frequency and pulse width, all but four of the programmable pacemakers presented sudden complete failure after different radiation doses. We conclude that direct pulse radiation at therapeutic levels of programmable pacemakers should be avoided.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.