Abstract

We have investigated the energy resolution and absolute photopeak efficiency of CZT detectors with three anode pixel geometries after three data correction steps. Before data corrections, the energy resolution is 4–5% at 122 keV, and the efficiency is 46–70% for a 21 keV photopeak window. Charge sharing events, which make up 10–25% of all events, are a significant cause of the poor efficiency. We discovered that large gaps between anode pixels have a strong negative impact on both charge sharing and absolute photopeak efficiency. Using three-dimensional position-dependent energy corrections, we were able to compensate for some of this reduced sensitivity, increasing the efficiency by up to 35% to a final efficiency of 62–80% of incoming gamma events.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.