Abstract

The in vitro toxicity of 15 different cobalt-chromium alloys used in removable partial dentures and dental implants was studied. The composition of the alloys was determined by X-ray spectrometry. Alloys were tested on human oral fibroblasts in low density, semiconfluent or confluent cultures. The cultures were observed in a phase contrast microscope over a period of 3 wk. The alloys could be divided into three groups depending upon the degree of cell injury evoked. Group I alloys caused no cell injury. Group II alloys caused slight to moderate cell injury seen first after 3-7 days of exposure. Group III alloys caused severe cell injury with signs of cell damage detected after 12 h and total cell death after 3-14 days. The alloys' composition varied considerably with regard to cobalt, chromium and nickel as well as other components. Nevertheless, group II and III alloys had a higher nickel content than group I alloys in general.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.