Abstract

A mass spectrometer array is described in which each array element is a cylindrical ion trap (CIT) within which an approximately quadrupolar, time-varying, field is established. The individual traps are of different sizes, so that when the array is operated with a fixed rf potential, ions of different masses (or mass ranges) are stored in each trap. By choosing the dimensions of each CIT element in the array, a multiple ion monitoring experiment can be performed. For example, in a two-element array with elements having internal radii of 5 and 4 mm, the smaller trap selects for m/z 91 and the larger for m/z 57, corresponding to characteristic aromatic and aliphatic hydrocarbon ions. Ion storage using both rf/dc (apex) isolation and the stored waveform inverse Fourier transform method is demonstrated.The array reduces the complexity of the electronics needed to operate the ion trap, which should make it suitable for use in a miniature mass spectrometer system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.