Abstract
A cylinder plate assay procedure was studied by 10 laboratories. For premix feeds, 3 samples of bacitracin methylene disalicylic acid and 3 samples of bacitracin zinc premixes covering the range of 10 to 50 g/lb were used. The repeatability standard deviation was 2.11, and the reproducibility standard deviation was 2.13. The average recovery of bacitracin was 101.5%. The method has been adopted official first action. For finished feeds, 6 samples of bacitracin methylene disalicylic acid and 6 samples of bacitracin zinc covering the range of 10 to 800 g/ton were used in the study. The procedure included a sample cleanup step using disposable reverse phase columns. This step appears to be the cause of the poor results reported by most collaborators. Continued study is needed to develop an acceptable method for finished feeds.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.