Abstract

To the Editor.—Interference from lipemia can significantly impede accurate analysis of a variety of laboratory tests independent of instrumentation or manufacturer.1 Conventional treatment of lipemia involves ultracentrifugation of specimens at approximately 199 000g to pellet and clarify interfering lipids. Although effective, this approach requires additional instrumentation, which may be costly for smaller, satellite laboratories offering a limited testing menu. Triaging lipemic samples to larger reference laboratories or core facilities may be impractical because of extended turnaround times and specimen rejection of lipemic samples, thus hindering patient care.Recently, our institution evaluated a lipid-precipitating, nonionic-cyclodextrin polymer, LipoClear (Statspin, Westwood, Massachusetts)2 as an alternative to ultracentrifugation in satellite laboratories as a means of reducing cost and improving turnaround time. Plasma and serum samples (n = 40) with grossly elevated lipemia (≥3+) were collected and refrigerated before analysis. Lipemia indices before treatment ranged from 3 to more than 10 (median = 6+). Specimens were split and treated to conventional ultracentrifugation (15 minutes at 199 000g) or with LipoClear (5 minute incubation of 500 μL of sample and 100 μL of LipoClear, followed by 7 minutes at 3461g). Clarified specimens were transferred to new tubes and analyzed on a Beckman Coulter DxC800 (Fullerton, California).LipoClear-treated specimens showed good agreement with paired ultracentrifuged specimens for all assays tested: sodium, potassium, chloride, carbon dioxide, calcium, glucose, blood urea nitrogen, total protein, albumin, alanine aminotransferase, alkaline phosphatase, aspartate aminotransferase, total bilirubin, phosphorous, creatinine, and creatine kinase. Linear regression analysis demonstrated acceptable performance across clinical decision points for all analytes. Correlation coefficients ranged from R = 0.92 to R = 0.99 for all analytes, except sodium and total protein, whose coefficients were R = 0.90. Sodium demonstrated a mean positive bias of 3.6 mEq/L (mmol/L) using LipoClear, compared with ultracentrifugation (Figure 1, A and B). This was considered within the limits of acceptability for the assigned total error budget of ±4 mEq/L (mmol/L) at pertinent medical decision points.3 Total protein demonstrated a mean negative bias of 0.65 (g/dL) using LipoClear, compared with ultracentrifugation. Previous work has compared preanalyte and postanalyte measurements on the Beckman Synchron LX-20, after artificial spiking of nonlipemic samples with the synthetic lipid Intralipid (Uppsala, Sweden).4 In this study, no discordance between recovery with ultracentrifugation and LipoClear was observed with sodium or total protein, highlighting potential limitations for Intralipid as a universal substitute for assessing endogenous lipemic interferences. Additionally, 78% of LipoClear-treated samples showed lower lipemia indices after treatment than did their paired ultracentrifuged specimens, suggesting a more-complete removal of interfering lipids. The results here are limited to performance on the DxC800, and it is the responsibility of the laboratory to validate any preanalytic specimen manipulation not covered in the manufacturer's instructions for use.Analytic performance aside, implementation of LipoClear in satellite laboratories reduced cost and drastically improved turnaround time for reporting lipemic specimens. Process mapping identified 11 steps necessary for dealing with lipemic specimens before chemotherapy infusion for our oncology clinic (Figure 2). Before any clinical procedure, an additional waiting time of at least 2 hours was not uncommon because of courier pickup, delivery, ultracentrifugation at the offsite laboratory, and result reporting. Incorporation of the onsite treatment of lipemia reduced the number of steps from 11 to 7 and added only 15 minutes compared with normal specimens. The economic barrier of approximately US $14,000 for purchasing ultracentrifuges for every satellite laboratory, as well as the need for a dedicated air supply for their operations, had previously prevented onsite treatment of affected specimens. Incorporation of this protocol returned results for lipemic specimens 87% faster and reduced cost 3-fold. Collectively, the elimination of costly equipment and courier fees, coupled with a shorter turnaround time, added value for both our satellite laboratories and for our patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.