Abstract
Spectrophotometric determinations of heavy metal ions with xylenol orange (XO) have been carried out using cyclic flow injection analysis. A cation-exchanger mini-column (Amberlite IR124) was incorporated after a flow-through cell for the on-line regeneration of the used reagents and the accumulation of heavy metal ions. A solution of 1.0×10−5 M XO in 0.01 M acetate buffer (pH 5.0) in a single reservoir (100 mL) was continuously circulated, and sample solutions were injected into the stream. Cadmium(II), copper(II), iron(III), lead(II), zinc(II) and lanthanum(III) were sensitively monitored and determined. The regeneration and recycling of the XO reagent allowed as many as 50 repetitive determinations of heavy metal ions with the same 100 mL circulating reagent solution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.