Abstract

The purpose of this study was to evaluate cyclic fatigue of .06 ProFile Ni-Ti rotary instruments after clinical use in molar teeth. In group 1, instruments size 40-15 were used in a crown-down technique using 2.5% NaOCl as an irrigant. Fifty-two molars were included and 13 sets of Profile Ni-Ti rotary instruments were used. Each set of instruments was used in four molars, and was steam autoclaved before each use. Group 2 (10 sets of new ProFile Ni-Ti rotary instruments) was the control group. Cyclic fatigue was tested by rotating the instruments in a 90 degrees metallic tube until they broke. One-way analysis of variance did not show any statistically significant differences amongst the files from both groups regarding cyclic fatigue. It was concluded that sterilization and clinical use in the presence of NaOCl did not lead to a decrease in the number of rotations to breakage of the files.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.