Abstract

Reciprocating motion expands the lifetime of endodontic instruments during the preparation of severely curved root canals. This study aimed to investigate the time to fracture (TTF) and number of cycles to failure (NCF) of different reciprocating instruments (n = 20 in each group) at body temperature using a dynamic testing model (amplitude = 3 mm). Reciproc Blue (RPB), size 25/.08, WaveOne Gold (WOG) 25/.07, Procodile (Proc) 25/.06, R-Motion (RM_06) 25/.06 and R-Motion (RM_04) 30/.04 instruments were tested in their specific reciprocating motion in artificial matching root canals (size of the instrument ± 0.02 mm; angle of curvature 60°, radius 5.0 mm, and centre of curvature 5.0 mm from apical endpoint). The number of fractured instruments, TTF, NCF, the and lengths of the fractured instruments were recorded and statistically analysed using the Chi-Square or Kruskal-Wallis test. Both TTF (median 720, 643, 562, 406, 254 s) and the NCF (3600, 3215, 2810, 2032, 1482 cycles) decreased in the following order RM_06 > RPB > RM_04 > Proc > WOG with partially significant differences. During testing, only six RM_06 instruments fractured, whereas 16/20 (RPB), 18/20 (Proc), and 20/20 (RM_04, WOG) fractures were recorded (p < 0.05). Within the limitations of the present study, blue-coloured RPB and RM instruments exhibited a significantly superior cyclic fatigue resistance compared to SE-NiTi and Gold-wire instruments. Heat treatment, cross-sectional design and core mass significantly influenced the longevity of reciprocating instruments in cyclic dynamic testing.