Abstract

This study evaluated the resistance to cyclic fatigue and the torsional strength of Prima One Gold® and WaveOne Gold® reciprocating systems. Cyclic fatigue was tested in an artificial canal with 60º curvature, a 5 mm radius of curvature at 22°C (n=10) and 36.5°C (n=10). The torsional strength test followed ISO 3630-1, measuring torque and deflection angle until fracture (n=10). The t-test was used to compare the two groups, and the significance level was set at 5%. The Prima One Gold group presented greater resistance to cyclic fatigue at room temperature (P=0.001) and body temperature (P<0.001). For the torsional resistance test, the Prima One Gold group had a lower maximum torque value (P=0.029), despite having a greater deflection angle (P<0.001). The Prima One Gold group showed greater resistance to cyclic fatigue regardless of temperature and a greater deflection angle. The WaveOne Gold group showed a greater maximum torque value than the Prima One Gold group.