Abstract

This case report of 74-year-old man with trigeminal neuralgia is presented to underscore the importance of evaluating the entire treatment plan, especially when delivering large doses where even a low percentage of the prescription dose can contribute a substantial dose to an unintended target. The patient was treated using the CyberKnife stereotactic radiosurgery system utilizing a nonisocentric beam treatment plan with a 5-mm fixed collimator generating 111 beams to deliver 6000 cGy to the 79% isodose line with a maximum dose of 7594 cGy to the target. Two weeks after treatment the patient's trigeminal neuralgia symptoms resolved; however, the patient developed oral mucositis due to the treatment. This case report reviews the cause of mucositis and makes recommendations on how to prevent unintended targets from receiving treatment.