Abstract

The response of a cyanide ion-selective electrode (ISE) in the presence of copper has been investigated. The interferences of metal ions, like copper, on the response of the cyanide ISE prevents its use for routine direct analysis of metal cyanide containing waste water. In this work an improved understanding of the chemistry at the electrode, and the effects of different cyanide and copper concentrations and matrix pH values on the electrode response, has been developed. Based on this insight, a semi-empirical approach is described to allow calibration of the ISE enabling the estimation of total cyanide. While not as robust as, for example, total cyanide distillation, it provides an inexpensive, rapid and convenient approach for more frequent and widespread screening of industrial waste waters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.