We demonstrate the utility of indotricarbocyanine (Cy7) conjugates of the phycobiliproteins phycoerythrin (PE) and allophycocyanin (APC) in flow cytometry. This is the first demonstration of the use of an APC tandem dye for fluorescence measurements. These resonance energy transfer tandem dyes can be excited by the phycobiliprotein-specific excitation wavelengths and fluoresce at wavelengths above 780 nm. The tandem dyes, when conjugated to antibodies, are suitable for flow cytometry and other immunofluorescence applications. These conjugates are easily detectable above the very low autofluorescence in this part of the spectrum. Indeed, the Cy7-conjugated PE tandem (Cy7PE) has a "brightness" (fluorescence signal over cellular autofluorescence) comparable to that of fluorescein, and the Cy7APC tandem has a "brightness" comparable to that of APC. These tandems are also easily distinguished from other commonly used fluorophores, making them suitable for high-order multiparametric analysis. We show an example of six-color immunofluorescence analysis by flow cytometry, simultaneously measuring fluorescences from fluorescein, PE, Cy5PE, Texas red, APC, and Cy7APC.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.