Abstract

This paper reports on the customized thinning of neural probes based on silicon (Si) using deep reactive ion etching (DRIE) as a post-processing step. The reduced probe dimensions are expected to minimize local tissue trauma, while guaranteeing probe integrity during implantation. For DRIE, the probes are partially masked by a micromachined Si cover chip comprising tailored cavities enabling any desired thinned length l and probe thickness t by a proper choice of cover chip design and DRIE parameters, respectively. A broad variety of probe designs were realized with shank tip thicknesses ranging from 35 µm down to 2 µm. All probes could successfully be implanted into a brain tissue phantom, demonstrating a pronounced reduction in insertion force from 0.55 mN for unprocessed probes to 0.08 mN for 2-µm-thin shanks. When the dura mater was mimicked by a polyethylene (PE) membrane, forces were reduced from 28.9 mN to 16.6 mN for 15-µm-thin shanks.