Abstract

A technique is presented for modifying the Blom‐Singer tracheostoma valve housing to improve peristomal skin adherence for hands‐free tracheoesophageal speech production following total laryngectomy. The finished product is thin, flexible, and maintains enhanced tear strength. The edges of a standard housing device for the Blom‐Singer valve are trimmed to allow for better accommodation to the tissue surface without distortion. A punch biopsy instrument is used to cut uniform holes around the housing device to allow for permeation of the silicone mixture for better mechanical adherence of the materials.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.