Abstract

Saturation curves for free-air ionization chambers with electrode gap widths of 4.2, 8.4 and 18 mm were obtained for 10 and 15 keV undulator synchrotron radiation thinned with a 230 Hz rotating-disk chopper. Ion recombination in free-air ionization chambers was found to be inversely proportional to the applied electric field, and an expression that satisfactorily reproduced the ion-recombination rate is determined. A comparison of the expressions for continuous and pulsed X-rays revealed that chopped high-intensity X-rays require a higher voltage to attain saturation when the product of the pulse width and electric field exceeds a value that depends on the X-ray energy. This behaviour was observed explicitly for 10 keV X-rays in measurements with the ionization chamber placed before and after the chopper.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.