Abstract

Abstract Co Self-Powered Neutron Detector (SPND) is confronted with the problem of material consumption, which causes the response current can neither reflect the change of neutron flux in time nor be proportional to the neutron flux. In this paper, a deconvolution-based method is established to solve this problem. First of all, a step signal of neutron flux is taken as an example to analyze its performance. When the material consumption of Co SPND is 10%, after compensation, the response current can be in correspondence of neutron flux. Finally, the effects of this model in different Signal-to-Noise Ratio are analyzed, which fully confirms the truth of its excellent performance for compensating Co SPND's signal.