Abstract

Eighteen patients underwent CT guided stereotactic biopsies using either a Riechert-Mundinger stereotactic frame modified for use in the CT scanner or the CT-based intracranial landmark technique of Gildenberg and a Todd-Wells frame. In seven patients undergoing biopsy with the modified frame technique, representative tissue samples were obtained in all cases. Eleven patients had biopsies using the Gildenberg technique and representative tissue samples were obtained in 10 of the 11. No major postoperative complications occurred in either group. Both techniques allowed samples of intracranial tissue to be obtained safely and efficiently. The modified frame technique provided a higher degree of accuracy while the Gildenberg technique required significantly less CT time and provided a satisfactory degree of accuracy. In seven patients with ring enhancing lesions, biopsies of the enhancing rim provided diagnostic tissue while biopsy of the central low density center of the lesion in three patients provided necrotic material which was useful in histologically grading the tumour.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.