Computed tomography automatic exposure control (CT-AEC) technique is calculated from a localizer radiograph. When we perform neck and chest CT examination, at first, we acquire localizer radiograph and neck images by placing the arm in a lowered position. Next, the arm is raised for the chest scan. Therefore, the localizer radiograph and subject information are different in the chest scan. In this situation, the chest scan with the use of the CT-AEC causes radiation over-dose. The purpose of this study is to optimize the CT-AEC by controlling noise index (NI), and make a chest CT scan condition considering the position of the arms. We measured the image noise (SD) in the phantom by using CT-AEC. In addition, dose length product (DLP) was recorded. Moreover, we examined the correlation with the clinical images. The results of our experiments show that radiation dose can be reduced with the image quality kept by controlling NI.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.