Abstract

This work aimed at assessing the doses delivered in Switzerland to paediatric patients during computed tomography (CT) examinations of the brain, chest and abdomen, and at establishing diagnostic reference levels (DRLs) for various age groups. Forms were sent to the ten centres performing CT on children, addressing the demographics, the indication and the scanning parameters: number of series, kilovoltage, tube current, rotation time, reconstruction slice thickness and pitch, volume CT dose index (CTDI(vol)) and dose length product (DLP). Per age group, the proposed DRLs for brain, chest and abdomen are, respectively, in terms of CTDI(vol): 20, 30, 40, 60 mGy; 5, 8, 10, 12 mGy; 7, 9, 13, 16 mGy; and in terms of DLP: 270, 420, 560, 1,000 mGy cm; 110, 200, 220, 460 mGy cm; 130, 300, 380, 500 mGy cm. An optimisation process should be initiated to reduce the spread in dose recorded in this study. A major element of this process should be the use of DRLs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.