Abstract

To evaluate the usefulness of the triple-lumen balloon catheter in the serial performance of CT arterial portography (CT-AP) and CT arteriography (CT-A). A combined CT-AP and CT-A examination of 58 patients was carried out in which a cobra-type triple-lumen balloon catheter was used. CT-AP was performed by injecting contrast medium either into the splenic artery through a side-hole in the catheter proximal to the balloon inflated in the common hepatic artery, or into the superior mesenteric artery through an end-hole in the catheter. Then CT-A was serially performed by delivering contrast medium either to the common hepatic artery or the proper hepatic artery from the end-hole, or to the accessory right hepatic artery through a side-hole proximal to the inflated balloon. Sufficient CT-APs were obtained in 53 of the 58 patients (91%), CT-A in 42 (72%), and both in 42 (72%). Incomplete CT-AP was due to technical failure or anatomical anomaly, as was incomplete CT-A. No complications were seen. The triple-lumen balloon catheter technique is useful and convenient in the serial performance of CT-AP and CT-A.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.