Abstract

CT angiography prior to endovascular aortic surgery is the standard non-invasive imaging method for evaluation of aortic dimensions and access sites. A detailed report is crucial to a proper planning. We assessed Artificial Intelligence (AI)-algorithm accuracy to measure vessels diameters at CT prior to transcatheter aortic valve implantation (TAVI). CT scans of 50 patients were included. Two Radiologists with experience in vascular imaging together manually assessed diameters at nine landmark positions according to the American Heart Association guidelines: 450 values were obtained. We implemented TOST (Two One-Sided Test) to determine whether the measurements were equivalent to the values obtained from the AI algorithm. When the equivalence bound was a range of ± 2mm the test showed equivalence for every point; if the range was equal to ± 1mm the two measurements were not equivalent in 6 points out of 9 (p-value > 0.05), close to the aortic valve. The time for automatic evaluation (average 1min 47s) was significantly lower compared with manual measurements (5min 41s) (p < 0.01). In conclusion, our results indicate that AI-algorithms can measure aortic diameters at CT prior to endovascular surgery with high accuracy. AI-assisted reporting promises high efficiency, reduced inter-reader variabilities and time saving. In order to perform optimal TAVI procedure planning aortic root analysis could be improved, including annulus dimensions.