Abstract
A liquid-nitrogen-cooled monochromator for high-energy X-rays consisting of two bent Si(111) Laue crystals adjusted to sequential Rowland conditions has been in operation for over two years at the SRI-CAT sector 1 undulator beamline of the Advanced Photon Source (APS). It delivers over ten times more flux than a flat-crystal monochromator does at high energies, without any increase in energy width (DeltaE/E approximately 10(-3)). Cryogenic cooling permits optimal flux, avoiding a sacrifice from the often employed alternative technique of filtration - a technique less effective at sources like the 7 GeV APS, where considerable heat loads can be deposited by high-energy photons, especially at closed undulator gaps. The fixed-offset geometry provides a fully tunable in-line monochromatic beam. In addition to presenting the optics performance, unique crystal design and stable bending mechanism for a cryogenically cooled crystal under high heat load, the bending radii adjustment procedures are described.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.