Abstract
We describe an apparatus for the cryogenic landing of particles from the ion beam of a mass spectrometer onto transmission electron microscope grids for cryo-electron microscopy. This system also allows for the controlled formation of thin films of amorphous ice on the grid surface. We demonstrate that as compared to room temperature landings, the use of this cryogenic landing device greatly improves the structural preservation of deposited protein-protein complexes. Furthermore, landing under cryogenic conditions can increase the diversity of particle orientations, allowing for improved 3D structural interpretation. We conclude that this approach allows for the direct coupling of mass spectrometry with cryo-electron microscopy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.